Model Number PCF-PH190L
Medical Device Problem Code Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.Date of event is unknown.Additional information will be provided if available.A definitive root cause could not be identified.Based on the results of the investigation, the most probable cause of the complaint was not established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer returned the colonovideoscope to the service center for a separate issue.During the device analysis, the service repair technician indicated white foreign materials in air/water channel and universal cord.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III COLONOVIDEOSCOPE
Common Device Name
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25545643
Report Number9610595-2026-49512
Device Sequence Number5571843
Product Code FDF
UDI-Device Identifier04953170305238
UDI-Public04953170305238
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K131780
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberPCF-PH190L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/15/2026
Initial Report FDA Received Date06/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/19/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-