| Model Number |
6487800 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On-site investigation was performed by our field service engineer.The claimed issue was not reproduced during troubleshooting.According to received information in service report, no part was replaced.The ventilator passed all functional and safety tests and was cleared for clinical use.Reported error indicates that the safety valve is detected as open without fulfilled opening conditions.A safety valve open alarm shall be triggered when a low level of the signal safety valve closed is detected for more than 8 sec and before 10 sec.Recommended action in case of technical error occurrence is to check inspiratory channel or safety valve pull magnet or expiratory channel board.Occurrence of reported error may lead to stop of ventilation.The reported issue was confirmed in provided device's log.The cause of the failure was not established.
|
| |
|
Event or Problem Description
|
|
It was reported that the ventilator generated a technical error code indicating an open safety valve.There was no patient harm.Manufacturer's ref.#: (b)(4).
|
| |
|