• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS INTELLIVUE MX40 2.4GHZ; MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS INTELLIVUE MX40 2.4GHZ; MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number MX40 - 2.4
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2026
Type of Reportable Event Malfunction
Event or Problem Description
Philips received a complaint on a mx40 2.4 ghz smart hopping device indicating that on 7 jan 2026, unit 575473 went offline for 20 seconds.The customer indicated there have been telemetry interruptions in the cardiologist's ward, examination ward and cardiac intensive care unit, where they have lost contact with the central unit (log "equipment offline") so that healthcare staff have not been able to adequately monitor patients.The customer expressed concern over interruptions in a cardiology department where continuous cardiac monitoring is necessary with patients admitted for investigation with cardiac monitoring or patients at risk of life-threatening arrhythmias.The customer further noted that the procedures for these interruptions are resource-intensive and risk other tasks being interrupted, not performed or delayed.The device was in use on a patient at time of event, there was no adverse event reported.
 
Additional Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX40 2.4GHZ
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
deborah currlin
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key25547454
Report Number1218950-2026-100425
Device Sequence Number5571896
Product Code MHX
UDI-Device Identifier00884838082243
UDI-Public(01)00884838082243
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K113125
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 06/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMX40 - 2.4
Device Catalogue Number865351
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2026
Initial Report FDA Received Date06/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/22/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-