| Model Number |
IN2110X15B |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information received from the device evaluation indicating an additional critical ventilator service required alarm had occurred and was documented in the device error log indicating the device software has detected a large pressure drop between the monitor and outlet pressure sensors (i.E.Across the machine flow sensor).
|
| |
|
Additional Manufacturer Narrative
|
|
The device mentioned in this complaint has exceeded its useful life per the applicable ifu.Therefore, the dhr for this device will not be reviewed at this time.However, the dhr files for this device are available and can be accessed upon specific request.The reported failure of the trilogy evo machine flow sensor is accommodated in the product risk management file as being linked to hazard with description patient exposure to function / deterioration of function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
|
| |
|
Event or Problem Description
|
|
The manufacturer received information that a trilogy evo ventilator was returned to a third-party service center for completion of preventative maintenance.There was no patient involvement, and no harm or injury reported.On evaluation, a critical device error code e-386 was noted in the error log indicating machine flow sensor offset greater than 0.2 slpm.The machine flow sensor required replacement to address this issue.Additional findings not related to the malfunction/issue: the system printed circuit board assembly required replacement and interior components were noted to be contaminated with dust including the machine flow sensor.Preventative maintenance with required component replacements required.
|
| |
|
Search Alerts/Recalls
|