| Model Number |
6372 |
| Medical Device Problem Code |
Low impedance (2285)
|
| Health Effect - Clinical Code |
Movement Disorder (4412)
|
| Date of Event |
06/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported the patient is experiencing ineffective therapy due to high and low impedances on the leads.Patient slightly bumped their head.Surgical intervention may take place at a later date to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 - concomitant medical products - 6662.Date of event is estimated.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d1 - brand name\
d4 - additional device information\
\
the report of ¿ineffective therapy¿ was not confirmed.Analysis of the returned extension (b) did not identify any anomalies or broken internal wires, and the extension passed continuity and output resistance testing.
|
| |
|
Event or Problem Description
|
|
Additional information received indicates the extensions were explanted and replaced to address the issue.Effective therapy was established.Implanted (b)(6) of 2024.Exact date of implant is unknown.
|
| |
|