• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH ENFIT CONNECTOR; DH CPK NG TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH ENFIT CONNECTOR; DH CPK NG TUBES Back to Search Results
Model Number 42-1225
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Avanos medical inc.Received a single report that referenced two different incidents, which were associated with separate units, involving two different patients.This is the first of two reports.Refer to 9611594-2026-00468 for the second report.It was reported that a nasogastric tube was occluded during initial use.
 
Additional Manufacturer Narrative
This event was initially deemed not reportable based on our reporting strategy active at the time of the alert date; however, following a recent retrospective review of complaints this event was made reportable out of an abundance of caution.The device history record for lot 30303736 was reviewed and the product was produced according to product specifications.The actual complaint product was returned for evaluation.The tube was decontaminated, the ports were flushed, and resistance was felt.The bolus end tip was observed.It appears to have some clear/ shiny matter blocking the bolus end opening.When the cut was created above the bolus component, it was observed a hardened material (shiny/ glossy appearance) stuck into the bolus opening.The root cause was determined to be related to the manufacturing process.All information reasonably known as of 18 jun 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as (b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH ENFIT CONNECTOR
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp 84048
MX   84048
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key25552628
Report Number9611594-2026-00467
Device Sequence Number10751592
Product Code KNT
UDI-Device Identifier00350770460734
UDI-Public(01)00350770460734(240)104607302(17)271010(10)30303736
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K821906
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number42-1225
Device Catalogue Number42-1225
Device Lot Number30303736
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2024
Initial Report FDA Received Date06/18/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-