Avanos medical inc.Received a single report that referenced two different incidents, which were associated with separate units, involving two different patients.This is the second of two reports.Refer to 9611594-2026-00474 for the first report refer to.It was reported that a nasogastric feeding tube clogged, and it broke in two when trying to unclog.The broken part was retrieved at the back of the throat.The patient had already been sedated and intubated at the time of occurrence.
|
|
This event was initially deemed not reportable based on our reporting strategy active at the time of the alert date; however, following a recent retrospective review of complaints this event was made reportable out of an abundance of caution.A review of the device history record is not possible as no lot number was provided; therefore, the udi-pi is unavailable.The actual complaint product was returned for evaluation.The tube was flushed through the y connector (proximal end) and was able to flush the solution through.The same goes on other part of tube (towards distal end).At the 84cm marked location, the tubing appeared to have expanded to form a balloon shape which burst in multiple direction causing a separation of the tube into two pieces.No blocking was observed.The root cause has been determined to be use error.All information reasonably known as of 18 jun 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|