|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
|
Additional/corrected data: d9, h3, h6.Device evaluation:
the device related to the reported event deflation was received on august 20, 2026, with lot number 3467193.Per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification (if applicable).Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
¿ deflation: observed an opening through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases, stress marks and wear abrasion are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|