| Model Number |
P1170G |
| Medical Device Problem Code |
Activation Problem (4042)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The baxter technician found the cpr handles needed to be replaced.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in mar 31, 2026.It is unknown if the facility performed any other preventative maintenance on this bed.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Examine the handles, brackets, and cables.Make sure that they are in good condition and secured to the bed.Make sure that the bracket and assembly are not bent or damaged.Make sure that the cable is not corroded.Replace or repair parts as necessary.Make sure that the screws are installed and fully tightened.Replace or tighten the screws as necessary.Fully raise the head section, then activate one of the cpr releases.Make sure that the head section lowers.Adjust the cpr cable as necessary.Do the same test on the other side of the bed.The technician replaced the cpr handles to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Event or Problem Description
|
|
Baxter received a report that careassist es155/255/455 had cpr function not activating.N issue.This occurred after patient use.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|