| Model Number |
72404011 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Nonstandard Device (1420); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404310 serial number: n/a batch/lot number: 1100757810 model/catalog description: pump ms unique identifier (udi) #: (b)(4).Model number/catalog number: 72404161 serial number: n/a batch/lot number: 1100812785 model/catalog description: reservoir 65ml pc unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Mechanical issue and hole/perforated are acknowledged within the instructions for use (ifu) and are not related to off-label use or failure to follow instructions.Based on the information available the cause that contributed to the reported event cannot be established; therefore, a conclusion code of cause not established was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with an inflatable penile prosthesis (ipp) presented with a nonfunctioning device.A revision surgery was performed, during which the physician confirmed a crack in the cylinder connecting tube.The device components were explanted and replaced, while the reservoir remained implanted and was not replaced.There were no patient complications.
|
| |
|