H3: product analysis: evaluation of the device determined that the reported malfunction of device seized was confirmed.The likely cause of failure could be detached distal bearing.It was also noted that the tube cannot extent/retract due to the rotating dial being stuck, tip of the tube was worn, input and output drive shaft was worn and laser markings were faded.G3: aware date used as date of analysis performed date as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|