Model Number DDU-A2300
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Troubleshooting of the aed with the customer identified that the aed pads were expired.During the aed's automated self-test, the aed would have identified that the pads were expired and attempted to alert the user by flashing its red asi and chirping.The aed would continue to chirp and flash until the aed's condition is addressed or the battery pack becomes completely depleted.The cause of the aed not powering on was related to a failure to set-up the aed and maintain the battery pack.Troubleshooting of the aed with the customer identified that once a new battery was installed and a manual self test run the aed functioned as designed and could stay in service.
 
Event or Problem Description
A customer reported that their aed's asi is blank, and the aed will not power on.The customer did not report this occurring during patient use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFELINE
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS
Manufacturer (Section D)
DEFIBTECH, L.L.C.
741 boston post rd.
suite 201
guilford CT 06437
Manufacturer (Section G)
DEFIBTECH, L.L.C.
14 commercial st.
branford
Manufacturer Contact
nissa soto
741 boston post road
suite 201
guilford, CT 06437
2034536654
MDR Report Key25562682
Report Number3003521780-2026-00949
Device Sequence Number21351475
Product Code MKJ
UDI-Device Identifier00815098020201
UDI-Public(01)00815098020201
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P160032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDDU-A2300
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2026
Initial Report FDA Received Date06/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/18/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-