U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENDOPATH; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENDOPATH; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ATS45
Medical Device Problem Codes Failure to Fire (2610); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 6/19/2026.D4: batch #827d32.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ats45 device was received with the firing trigger broken and it was returned inside of a plastic bag, also a tr45w reload was loaded in the device.The reload lockout spring was damaged, and the returned reload was partially fired 1/2.No functional test could be performed due to the condition of the device.Additionally the device was disassembled to verify the condition of the internal components and no damages were observed.The event reported was confirmed and it is related to improper use of the device.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If the re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number of 720d55 / 09ats45 , and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 827d32, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unspecified surgical procedure, it was observed that the device could not be fired and the device made a strange clicking sound.Changed to a new one to complete the procedure.There was no patient consequence nor surgical delay reported.No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
*   00969
3035526892
MDR Report Key25565771
Report Number3005075853-2026-04767
Device Sequence Number9996021
Product Code GDW
UDI-Device Identifier10705036001072
UDI-Public(01)10705036001072(17)300731(10)720D55
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K070887
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberATS45
Device Lot Number720D55
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2026
Initial Report FDA Received Date06/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-