• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION Autotome RX 44; Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION Autotome RX 44; Unit, Electrosurgical, Endoscopic (With Or Without Accessories) Back to Search Results
Model Number M00545170
Medical Device Problem Codes Break (1069); Failure to Conduct (1114)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.
 
Event or Problem Description
It was reported to boston scientific corporation that an autotome rx 44 was used in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure to treat bile duct stones performed on (b)(6) 2026.During the procedure, the device functioned normally on its first use.However, there was no response when it was activated a second time, and repeated attempts produced the same result.The autotome was then withdrawn, and examination revealed that its tip had fractured.The procedure was completed with another of the same device.No further information has been obtained despite good faith efforts.The reported event may suggest that the cutting wire broke.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Autotome RX 44
Common Device Name
Unit, Electrosurgical, Endoscopic (With Or Without Accessories)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
2546 calle primera
alajuela,
CS  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough,, MA 01752
EI   01752
9145716918
MDR Report Key25566219
Report Number3005099803-2026-02531
Device Sequence Number21331084
Product Code KNS
UDI-Device Identifier08714729444749
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K013153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00545170
Device Catalogue NumberM00545170
Device Lot Number0037872139
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2026
Supplement Date Received by Manufacturer08/06/2026
Initial Report FDA Received Date06/19/2026
Supplement Report FDA Received Date09/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
Patient Weight165 LB
-
-