• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA III BRONCHOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA III BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-XT190
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/17/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.The root cause could not be identified.Based on the results of the investigation, it is likely the following led to malfunction: environmental temperature problem resulted in component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
The customer returned the bronchovideoscope, which is an olympus asset with no reported complaint.During inspection it was found that the distal end plastic cover.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key25566336
Report Number9610595-2026-50138
Device Sequence Number9995626
Product Code EOQ
UDI-Device Identifier04953170402470
UDI-Public04953170402470
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K250862
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberBF-XT190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/01/2026
Initial Report FDA Received Date06/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-