(b)(4).Date sent: 6/19/2026.D4: batch#: 773d12.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ec60a device was returned with the jaws and closure trigger in the closed position.Also, the knife was noted partially advanced, the red manual knife reverse button was activated and the knife returned to the home position as expected and with a cecr60w partially fired 1/16 reload loaded on the device.The returned device and reload were tested for functionality in the articulated position by resetting and reloading it into the device.The device achieved its complete stroke firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.It is possible that while loading the reload, the reload was pushed farther back than the reload alignment stop windows resulting in the knife pushing the one-piece sled forward and locking the reload.Please reference the instruction for use for more information.Also, the condition of knife advanced should be noted that when using the reverse button ensure that the knife is fully back on its home position, if the knife remains advanced the device jaws would not open.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number of 791d80 / 23ec60a, and no non-conformance's were identified.A manufacturing record evaluation was performed for the finished device batch number 773d12, and no non-conformance's were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|