| Model Number |
DDU-A2450 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer initially reported a service code, review of the aed's internal log files determined that the service code after battery replacement was due to the previous battery fully depleting within the unit.Analysis of the aed's log files identified that the customer's failure to promptly address red asi warnings reduced battery life expectancy.On 12/17/25 the aed identified that the aed battery was getting low and attempted to alert the user by flashing its red asi and chirping.The aed continued to flash and chirp until 1/29/26 when the battery pack became completely depleted.There was no evidence in the electronic logs of any human interaction to address the aeds condition until 3/24/26 when a replacement battery pack was inserted into the aed and the aed powered on.
|
| |
|
Event or Problem Description
|
|
A customer reported that upon inserting a new battery pack, their aed powered on and prompted a service code.The customer did not report this occurring during patient use.
|
| |
|