• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 INHANCE; SHOULDER IMPLANT - GLENOSPHERE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 INHANCE; SHOULDER IMPLANT - GLENOSPHERE Back to Search Results
Catalog Number 550532004
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Fall (1848); Joint Dislocation (2374)
Date of Event 05/28/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that implants were mal-positioned, possibly from a patient fall.This caused patient to keep dis-locating.Full revision was required.Affected side: right.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INHANCE
Common Device Name
SHOULDER IMPLANT - GLENOSPHERE
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key25576440
Report Number1818910-2026-10576
Device Sequence Number17431571
Product Code PHX
UDI-Device Identifier10603295546665
UDI-Public(01)10603295546665(17)300731(10)354927
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K212737
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number550532004
Device Lot Number354927
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/28/2026
Supplement Date Received by Manufacturer07/03/2026
Initial Report FDA Received Date06/22/2026
Supplement Report FDA Received Date07/03/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-