ALCON GRIESHABER AG DIATHERMY PROBE DSP; UNIT, PHACOFRAGMENTATION, ACCESSORY, PROBE Back to Search Results
Catalog Number 339.31
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/28/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned instrument was received in its tyvek pouch, showing the lot information.The instrument was visually inspected and showed no evident abnormality.The electrical resistance measurements indicate that there was insufficient contact, leading to the instrument not functioning.When the instrument was opened to visually inspect the inner contacts, some of them were unintentionally destroyed (destructive instrument testing).No other defects or deficiencies could be identified on the contacts after opening the instrument.As the pouch was opened by the customer, the product was handled.Therefore, the point in time when the malfunction occurred can no longer be determined, nor can the strain put on the device be reconstructed.The customer¿s complaint is confirmed, but a root cause for the reported issue cannot be determined.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Non-conformance report provided by originator.A non-conformance-based review of the lot number was performed and did not reveal any potential contributing factors to the reported complaint.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.The complaint data will continue to be monitored for evidence of adverse trending and further action will be taken, as appropriate.No further action warranted at this time.No root cause for the customers reported event could be determined, since the situation that led to the instrument failure cannot be reconstructed.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during surgery, the ophthalmic probe had a poor connection.The procedure was successfully completed on the same day.Patient impact has not been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIATHERMY PROBE DSP
Common Device Name
UNIT, PHACOFRAGMENTATION, ACCESSORY, PROBE
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key25577213
Report Number3003398873-2026-00123
Device Sequence Number15719549
Product Code HQC
UDI-Device Identifier07612717071384
UDI-Public(01)07612717071384(17)280531(11)250619(10)17H65L
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063583
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number339.31
Device Lot Number17H65L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/29/2026
Initial Report FDA Received Date06/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/19/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-