SYNTHES GMBH TFN-ADVANCE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 04.037.043S
Medical Device Problem Code No Apparent Adverse Event (3189)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/24/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: e1 initial reporter address line 1: (b)(6).H3, h6: part:04.037.043s lot: 68564p4 manufacturing site: werk selzach supplier: (b)(4) release to warehouse date: 22 may 2025 expiration date; 01 may 2035.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed distal part of the prong that locks the blade into place is broken, remnants were not observed on the photo.With the information provided is not possible to determine a potential cause at this moment.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.Any conclusions from the investigational input provided have to be placed into context with all other relevant factors as the device was not returned, and as-received condition could not be assessed, a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the tfna fem nail ø10 130° l200 timo15 would have contributed to the device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Event or Problem Description
It was reported that there was a cut-through of the helix blade, with broken set screw of the tfna.The patient underwent revision surgery.Implants were destroyed by the facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFN-ADVANCE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key25578627
Report Number8030965-2026-06413
Device Sequence Number5166244
Product Code HSB
UDI-Device Identifier10886982095992
UDI-Public(01)10886982095992(17)350501(10)68564P4
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.037.043S
Device Lot Number68564P4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/19/2026
Initial Report FDA Received Date06/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-