|
H3: product analysis :evaluation of the device confirmed broken device.The most likely cause of failure was detached distal bearing.It was also noted that the tube was deformed and the laser markings were faded.B3: date of notification: 2026-04-24 (date of event or estimated date of incident not provided to manufacturer).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|