• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION MUSTANG?; STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION MUSTANG?; STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS Back to Search Results
Model Number H74939171060470
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D2b - pro code (product code): fge, lit.G4 - premarket / 510(k): k141521, k141597.Investigation results: device technical analysis: upon receipt at our post market quality assurance laboratory, a visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The rated burst pressure for this device is 24 atmospheres as per mustang specification.The returned device was attached to an inflation unit.Positive pressure was applied when di water was observed to be leaking from a balloon pinhole located approximately 1mm proximal from the proximal markerband.The investigator was unable to fully inflate the balloon due to the pinhole leak in the balloon.A visual examination of the markerbands identified no issues.A visual and tactile examination identified no kinks or damage to the shaft and to the tip.No other issues were identified during the product analysis.Device history record (dhr) review it was confirmed this device met manufacturing specifications prior to distribution and there were no manufacturing deviations that could have contributed to the reported event.Labeling review review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review: a review of the mustang device confirmed that the reported event is a known event defined in the product's risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to patient condition.
 
Event or Problem Description
Reportable based on device analysis completed on 04jun2026.It was reported that balloon failed to inflate.The severely stenosed target lesion was located in the severely tortuous and moderately calcified arteriovenous fistula.A 6.0 x 40, 75cm mustang balloon catheter was advanced for dilation.During the third inflation at 23 atmospheres, the balloon did not fully expand well in the lesion.The procedure was completed with a different device.There were no patient complications as a result of this event.However, device analysis revealed a balloon pinhole located approximately 1mm proximal from the proximal markerband.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MUSTANG?
Common Device Name
STENTS, DRAINS AND DILATORS FOR THE BILIARY DUCTS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key25580129
Report Number2124215-2026-33077
Device Sequence Number13016801
Product Code FGE
UDI-Device Identifier08714729793335
UDI-Public(01)08714729793335(17)260823(10)32284977
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K141521
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2026
Device Model NumberH74939171060470
Device Catalogue NumberH74939171060470
Device Lot Number0032284977
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2026
Initial Report FDA Received Date06/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age59 YR
Patient SexFemale
-
-