| Model Number |
H74939171060470 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D2b - pro code (product code): fge, lit.G4 - premarket / 510(k): k141521, k141597.Investigation results: device technical analysis: upon receipt at our post market quality assurance laboratory, a visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The rated burst pressure for this device is 24 atmospheres as per mustang specification.The returned device was attached to an inflation unit.Positive pressure was applied when di water was observed to be leaking from a balloon pinhole located approximately 1mm proximal from the proximal markerband.The investigator was unable to fully inflate the balloon due to the pinhole leak in the balloon.A visual examination of the markerbands identified no issues.A visual and tactile examination identified no kinks or damage to the shaft and to the tip.No other issues were identified during the product analysis.Device history record (dhr) review it was confirmed this device met manufacturing specifications prior to distribution and there were no manufacturing deviations that could have contributed to the reported event.Labeling review review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Risk review: a review of the mustang device confirmed that the reported event is a known event defined in the product's risk management documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to patient condition.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 04jun2026.It was reported that balloon failed to inflate.The severely stenosed target lesion was located in the severely tortuous and moderately calcified arteriovenous fistula.A 6.0 x 40, 75cm mustang balloon catheter was advanced for dilation.During the third inflation at 23 atmospheres, the balloon did not fully expand well in the lesion.The procedure was completed with a different device.There were no patient complications as a result of this event.However, device analysis revealed a balloon pinhole located approximately 1mm proximal from the proximal markerband.
|
| |
|
Search Alerts/Recalls
|