• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (IRVINE) XEN 45 GTS; IMPLANT, EYE VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (IRVINE) XEN 45 GTS; IMPLANT, EYE VALVE Back to Search Results
Catalog Number 5513-001
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Further information regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event is a known potential adverse event addressed in the product labeling.
 
Event or Problem Description
Healthcare professional reported a xen®45 gts \"slider malfunction.\" device did not make patient contact.Surgery was completed with a backup device.
 
Additional Manufacturer Narrative
Additional, corrected, and/or changed data: d9, h3, h6\ \ device analysis: the sample cannot be evaluated for the category/subcategory of injector ¿ slider resistance since the injector was received with the slider knob detached from the injector and a functionality test could not be performed.\ \ the category/subcategory of injector ¿ slider resistance is unable to verify because the injector was received with the slider knob detached from the injector and a functionality test could not be performed.\ \ the complainant did not report detachment of the slider knob from the injector or damage to the window of the track.The condition of the slider knob and the window of the track is likely due to handling.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XEN 45 GTS
Common Device Name
IMPLANT, EYE VALVE
Manufacturer (Section D)
ALLERGAN (IRVINE)
2525 dupont drive
irvine CA 92612
Manufacturer (Section G)
ALLERGAN (IRVINE)
2525 dupont drive
irvine CA 92612
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key25586061
Report Number3011299751-2026-00067
Device Sequence Number12107790
Product Code KYF
UDI-Device Identifier10888628032439
UDI-Public(01)10888628032439(17)280831(10)65519
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K161457
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5513-001
Device Lot Number65519
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/17/2026
Initial Report FDA Received Date06/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-