| Catalog Number |
5513-001 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Further information regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a xen®45 gts \"slider malfunction.\" device did not make patient contact.Surgery was completed with a backup device.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, corrected, and/or changed data: d9, h3, h6\
\
device analysis: the sample cannot be evaluated for the category/subcategory of injector ¿ slider resistance since the injector was received with the slider knob detached from the injector and a functionality test could not be performed.\
\
the category/subcategory of injector ¿ slider resistance is unable to verify because the injector was received with the slider knob detached from the injector and a functionality test could not be performed.\
\
the complainant did not report detachment of the slider knob from the injector or damage to the window of the track.The condition of the slider knob and the window of the track is likely due to handling.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|