• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. GARBIN EVO, LINDE; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. GARBIN EVO, LINDE; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number LD2110X23B
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the dhr for this device, serial number h26166239d98d, did not find any non-conformances or abnormalities related to the reportable malfunction/ allegation identified in this complaint record during testing of the device prior to distribution.The reported failure of the trilogy evo machine flow sensor is accommodated in the product risk management file as being linked to hazard with description patient exposure to function / deterioration of function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.
 
Event or Problem Description
The manufacturer received information that a trilogy evo ventilator was returned to a third-party service center for completion of preventative maintenance.There was no patient involvement, and no harm or injury reported.On evaluation, a critical device error code e-386 was noted in the error log indicating machine flow sensor offset greater than 0.2 slpm.The machine flow sensor required replacement to address this issue.Additional findings not related to the malfunction/issue: the system printed circuit board assembly required replacement and interior components were noted to be contaminated with dust.Preventative maintenance with required component replacements required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GARBIN EVO, LINDE
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key25592721
Report Number2518422-2026-025023
Device Sequence Number17567855
Product Code CBK
UDI-Device Identifier00606959057432
UDI-Public(01)00606959057432
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberLD2110X23B
Device Catalogue NumberLD2110X23B
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date06/23/2026
Supplement Report FDA Received Date06/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/25/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-