| Model Number |
Mi1000 MED-EL CONCERT |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user was scheduled for a revision surgery due to 5-6 extra-cochlear channels.The user has been re-implanted.
|
| |
|
Event or Problem Description
|
|
The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: according to the information received from the field the recipient underwent a revision surgery due to post-operative migration of the active electrode out of cochlea.During revision damage to the active electrode array was seen and the device was explanted.Additionally, damage to the reference electrode caused by an external mechanical force acting on the reference protector was observed during the investigation.Other damages found during investigation are most likely related due to explantation.This is a final report.
|
| |
|