• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.710
Medical Device Problem Codes Output below Specifications (3004); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11: additional narrative: device evaluation: the actual device was returned for evaluation.The device was evaluated and the reported condition that the device¿s engine was weak was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device had a sticky trigger.The assignable root cause of this condition was determined to be traced to component failure due to wear.
 
Event or Problem Description
It was reported that during service and evaluation, it was determined that the battery oscillator device had a sticky trigger, component damage, and would not run.It was further determined that the device failed pretest for general condition, check for sticky triggers, and check oscillation frequency with frequency meter.It was noted in the service order that the device had a weak engine.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported; however, it was reported that the event occurred in (b)(6) 2026.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY OSCILLATOR II FOR BPL II
Common Device Name
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
elba bello
eimattstrasse 3
oberdorf 04436
SZ   04436
9083863534
MDR Report Key25597718
Report Number8030965-2026-06483
Device Sequence Number5405609
Product Code GEY
UDI-Device Identifier07611819491847
UDI-Public(01)07611819491847
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number530.710
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2026
Initial Report FDA Received Date06/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/07/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-