| Model Number |
BR1054096 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported failure of the speakers is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.
|
| |
|
Event or Problem Description
|
|
A trilogy 100 ventilator device was returned to the third-party service center for service.There was no allegation of patient harm.The device was not reported to be in patient use.During the evaluation, r ventilator service required error codes e106 (speaker 1 failure) and e107 (speaker 2 failure) were identified in the error log.Therefore, both speakers had failed.No documentation of a replaced component to address the issue.Additionally, no visible foam particles were visualized.The device log identified additional r ventilator service required error codes e147 (err_drift_prox_pressure_too_high) and e197 (err_int_li_ion_not_present).A final report is being submitted.If new information becomes available, a follow up/supplemental report will be completed.
|
| |
|
Search Alerts/Recalls
|