| Model Number |
TB-0535FCS |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, the investigation findings do not lead to a clear conclusion about the cause of the reported adverse event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the probe at the proximal tip of the thunderbeat device broke off.There was no patient involvement.
|
| |
|