| Catalog Number |
05.001.201 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11: additional narrative: device evaluation: the actual device was returned for evaluation.The device was evaluated and the reported condition that the switch on the device was getting stuck was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device had a sticky trigger.The assignable root cause of this condition was determined to be traced to component failure due to wear.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the battery handpiece device had a sticky trigger, the device failed the leak tightness test and would not run.It was further determined that the device failed pretest for leakage test using bubble emission technique, check for mechanical free moving, and check for sticky triggers, and check general function of device.It was noted in the service order that the switch was getting stuck.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported; however, it was reported that the event occurred in 2026.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|