(b)(6).Method: the subject device, ojr418 optiflow junior 2 nasal cannula was discarded and not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the information provided by the distributor, and our knowledge of the product.Results: the distributor reported that the tubing of an ojr418 optiflow junior 2 nasal cannula was detached from the cannula end during use.Conclusion: without the subject device being returned for evaluation, we could not determine the cause of the reported event.As part of the manufacturing process, all optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails specifications is discarded.In addition, representative samples from each batch are functionally tested to assess retention strength between the assembled components.If there are any failures the entire batch is quarantined for further investigation.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 nasal cannula.They also state the following: - "appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death)." - "ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient." - "do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).
|