Catalog Number 00515048201
Medical Device Problem Codes Mechanical Problem (1384); Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event is recorded by zimmer biomet under (b)(4).G2: foreign - event occurred in japan.Visual examination of the returned product identified device would not power on with the original battery pack but did power on and function normally in both modes when connected to a lab battery pack.Visual inspection of the interior of the pack found the batteries had leaked electrolyte inside of the pack.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.The root cause is attributed to a manufacturing issue exacerbated by a design issue.Able to confirm reported event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that during surgery the device did not work even the motor at all, and the battery pack of this complaint product was getting hot.Investigation found the batteries had leaked electrolyte inside of the pack.There was no patient injury, or delay.Due diligence is complete and there is no additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIP KIT
Common Device Name
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key25608824
Report Number0001526350-2026-00722
Device Sequence Number15903146
Product Code FQH
UDI-Device Identifier00889024375192
UDI-Public(01)00889024375192(17)270910(10)79272712
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number00515048201
Device Lot Number79272712
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/11/2026
Initial Report FDA Received Date06/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-