MAILLEFER INSTRUMENTS HOLDING SARL CARB.B.CAV.CONE SQUARE FG 012; BUR, DENTAL Back to Search Results
Catalog Number E013834101200
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 05/28/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since this event resulted in medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, it must be presumed that the malfunction would be likely to cause or contribute to a serious injury should it recur.As such, this event meets the definition of a reportable event per 21 cfr part 803.Involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused bur is available for evaluation.Nothing unusual to report was found during dhr review (batch #1713252).Root causes are not identified.We will track this kind of event and monitor the trend.Potential root causes may be incorrect technique, overuse (number of uses not communicated), excessive wear, patient condition and behaviour during treatment or material issue (no analysis of the broken fragments possible).In addition, there are further circumstances on the conditions in dentistry (used disinfectants, training/knowledge status), or any other environmental conditions, which are unknown to us and may also have an impact on the reported failure mode.Root causes are not identified.All complaints are monitored through the monthly product surveillance committee and a corrective action could be determined by the committee.
 
Event or Problem Description
In this event it is reported that a carb.B.Cav.Cone square fg 012 bur broke during use and collapsed onto the patient's tongue, causing the patient to bleed.The doctor applied pressure hemostasis on the patient and performed follow up treatment after the patient had no discomfort.No injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARB.B.CAV.CONE SQUARE FG 012
Common Device Name
BUR, DENTAL
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key25609776
Report Number8031010-2026-00087
Device Sequence Number18634915
Product Code EJL
UDI-Device IdentifierJ003E0138341012001
UDI-PublicJ003E0138341012001
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 06/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberE013834101200
Device Lot Number1713252
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date06/12/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2026
Initial Report FDA Received Date06/24/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-