Model Number 694758
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Codes Presyncope (4410); Syncope/Fainting (4411); Asystole (4442)
Date of Event 06/03/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continuation of d10: competitor lead implanted: on (b)(6) 2007, fr995-21 tissue valve implanted: on (b)(6) 2014.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead exhibited cross talk with cross chamber oversensing and occasional pacing inhibition.The patient experienced pre-syncope, syncope, and asystole.The rv lead was explanted and replaced.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE MRI SURESCAN
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key25611777
Report Number2649622-2026-18825
Device Sequence Number5520852
Product Code LWS
UDI-Device Identifier00613994249876
UDI-Public(01)00613994249876(17)120929
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/29/2012
Device Model Number694758
Device Catalogue Number694758
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2026
Initial Report FDA Received Date06/24/2026
Date Device Manufactured10/11/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DDBB1D1 ICD.
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-