(b)(4) date sent: 6/24/2026 d4: batch #958c03.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pse60a device was returned with the handle shrouds partially opened, and with two cecr60b reloads (b and c) present.Both reloads was received partially fired 1/3.The returned device and reloads were tested for functionality in the straight position by resetting and reloading them into the device.The device achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.The partially fired is consistent with an incomplete or interrupted cycle.It is possible that the damage on the handle shrouds was due to improper handling of the device.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished #pse60a, with lot/batch #c9ev6p, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 958c03, and no non-conformances related to the reported complaint condition were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|