(b)(4).Date sent: 6/24/2026.D4: batch #391c10.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one gst60g reload was received unfired, with no apparent damage, and with an opened package.The batch number on the reload showed that the reload was expired as of apr/30/2026.While we have no reason to believe this contributed to the reported event, it should be noted that the device returned appears to have been used after the expiration date.The returned reload was loaded into a test device.The returned reload was tested for functionality with a test device in the straight position and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.The event described could not be confirmed as the reload performed without any difficulties noted.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number of 441c98 / 22gst60g , and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 391c10, and no non-conformances related to the reported complaint condition were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|