| Model Number |
3186 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/07/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that following an unrelated surgical procedure wherein the device was not placed into surgery mode, the ipg became unable to communicate with external devices.Troubleshooting has been unable to resolve the issue.The bilateral leads had high impedances, and the patient was unable to enter their device into mri mode.Surgical intervention was undertaken wherein the system was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
The date of event is estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|