• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO VILLALBA SPRINT QUATTRO SECURE S MRI SURESCAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO VILLALBA SPRINT QUATTRO SECURE S MRI SURESCAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 6935M62
Medical Device Problem Codes Signal Artifact/Noise (1036); Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was further reported that the patient is being paced by the right atrial (ra) lead and the rv lead was never used to pace the patient.
 
Event or Problem Description
It was reported that the right ventricular (rv) lead displayed oversensing of noise at rates in the ventricular fibrillation (vf) zone.Defibrillation therapies were programmed off, and the lead remains in use.The patient is a participant in a clinical study.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: 407652 lead implanted: (b)(6) 2014.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S MRI SURESCAN
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key25613687
Report Number2649622-2026-18854
Device Sequence Number10218009
Product Code LWS
UDI-Device Identifier00613994718853
UDI-Public(01)00613994718853(17)150814
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/14/2015
Device Model Number6935M62
Device Catalogue Number6935M62
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2026
Supplement Date Received by Manufacturer06/24/2026
Initial Report FDA Received Date06/24/2026
Supplement Report FDA Received Date07/13/2026
Date Device Manufactured08/27/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DDPA2D4 ICD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-