| Catalog Number |
00515048201 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).Review found that the device would not power on.Visual inspection of the interior of the pack found the batteries had leaked electrolyte inside of the pack review of the device history record(s) identified no deviations or anomalies during manufacturing.The root cause is attributed to a manufacturing issue exacerbated by a design issue.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor trends.G2: japan.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, the unit did not work at all from the start of use.Battery expulsion was confirmed after final assessment.There was no patient impact and it is unknown if a delay occurred.Due diligence is complete.
|
| |
|