| Model Number |
720157-01 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Urinary Incontinence (4572)
|
| Date of Event |
05/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 114630016.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: ((b)(4) model number/catalog number: 72404127.Serial number: n/a.Batch/lot number: 113346010.Model/catalog description: control pump fgs iz.Unique identifier (udi) #: (b)(4).The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.There was no report of a device performance allegation.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.The reported patient symptoms of urinary incontinence are known risks associated with this device type and indicated as such in the instructions for use.A risk review was completed and confirmed that the event of urinary incontinence and mechanical issue were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available, a conclusion code of cause not established was assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the artificial urinary sphincter (aus) presented with recurring incontinence and that the device was no longer functioning.A surgical procedure was performed, and the device was removed and replaced.No further patient complications have been reported.
|
| |
|
Search Alerts/Recalls
|