| Catalog Number |
SCX-580 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
02/01/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
|
| |
|
Event or Problem Description
|
|
Patient reported "capsular contracture baker grade unknown", "hardness" and "higher up than the other".Patient later reported "baker iii" for the capsular contracture.Healthcare professional reported "baker grade iii capsular contracture" and "tighter, higher, and at times painful".This record is for the left side.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: h6.
|
| |
|
Event or Problem Description
|
|
Patient reported "capsular contracture baker grade unknown", "hardness" and "higher up than the other".Patient later reported "baker iii" for the capsular contracture.Healthcare professional reported "baker grade iii capsular contracture" and "tighter, higher, and at times painful".This record is for the left side.The device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, b7, d6b, h.6.
|
| |
|
Event or Problem Description
|
|
Patient reported "capsular contracture baker grade unknown", "hardness" and "higher up than the other".Patient later reported "baker iii" for the capsular contracture.Healthcare professional reported "baker grade iii capsular contracture" and "tighter, higher, and at times painful".Patient later reported "device was found to be ruptured" during explant surgery.This record is for the left side.The device was explanted.
|
| |
|
Search Alerts/Recalls
|