| Model Number |
6882000 |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On-site investigation was performed by our field service engineer.The claimed issue was not reproduced during troubleshooting.According to received information in service report, no parts were replaced to solve the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.Alarm o2 concentration low is activated when measured o2 concentration exceeds the set value by more than 5 vol.% below the set concentration value.The alarm is delayed 40 seconds after changing the o2 concentration setting.There is also an absolute minimum alarm limit of 18% o2 which is independent of operating settings.There can be two reasons for this alarm: gas delivery error and measurement error.Gas delivery error is when wrong gas concentration is delivered from the system, but the gas concentration is measured correctly.Measurement error is when correct gas concentration is delivered from the system, but the gas concentration is measured incorrectly.Incorrect o2 concentration delivered to the patient may in some cases indicating that the ventilation have been insufficient due to gas delivery error, which represent a reportable event.The reported issue was confirmed in provided device's logs.The cause of the reported event has not been determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the ventilator generated an alarm indicating low o2 concentration.There was no patient harm.Manufacturer's ref.#: (b)(4).
|
| |
|
Search Alerts/Recalls
|