| Model Number |
305C223 |
| Medical Device Problem Codes |
Gradient Increase (1270); Device Stenosis (4066)
|
| Health Effect - Clinical Code |
Aortic Valve Stenosis (1717)
|
| Date of Event |
04/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that 14 years and 6 months post implant of this 23mm aortic bioprosthetic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement.The reason for evaluation was reported as valve stenosis and high gradient of 65mmhg.It was reported that an intervention is required and on (b)(6) 2026 the patient successfully had a transcatheter valve-in-valve replacement.No additional adverse patient effects were reported.
|
| |
|