This event was initially deemed not reportable based on our reporting strategy active at the time of the alert date; however, following a recent retrospective review of complaints this event was made reportable out of an abundance of caution.The returned sample was evaluated and no visible tears, cuts, holes, or leakage could be identified.Leak testing with dyed water under pressure did not reproduce the reported condition.A review of the device history record is not possible as a valid lot number was not provided; therefore, the udi-pi is unavailable.Root cause could not be determined.All information reasonably known as of 24 jun 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as (b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
The customer reported leakage of enteral nutrition from the middle portion of a corflo feeding tube during use.Prior to use, resistance was encountered while flushing water through the tube; however, the obstruction appeared to clear, and the tube continued to be used.Subsequently, leakage was observed from the middle section of the tube.No patient injury was reported.
|