| Model Number |
DS2110X11B |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported failure of machine flow sensor drift above the specification (>0.2lpm) is accommodated in the product risk management file as being linked to hazard with description patient exposure to function / deterioration of function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review was performed for the complaint device serial number, (b)(6) review confirms that the device met the final acceptance criteria and was released for final distribution.
|
| |
|
Event or Problem Description
|
|
The manufacturer received a report that a trilogy evo ventilator was returned for service.No patient harm or injury was reported.The device was evaluated at a third-party service center, and the customer complaint was confirmed.Investigation identified a "vent inop" alarm due to the machine flow sensor drift exceeding specification limits (>0.2lpm).The flow sensor required replacement.No repair was performed.The device was subsequently destroyed and disposed of.
|
| |
|
Search Alerts/Recalls
|