• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR; FOLDERS AND INJECTORS , INRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR; FOLDERS AND INJECTORS , INRAOCULAR LENS (IOL) Back to Search Results
Model Number IV
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
A nurse reported that during cataract surgery with an intraocular lens (iol) implant procedure, the lens was not used due to an injector override, upon inspection they noticed that the iol became stuck in the cartridge.The procedure was completed on the same day after switching out the iol, cartridge, and injector.The iol did not enter the eye, as the issue was noticed during loading.The doctor believes the issue was caused by a faulty inserter.Additional information had been requested and received stating that there was patient contact, and it was due to bad inserter.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR
Common Device Name
FOLDERS AND INJECTORS , INRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key25620282
Report Number2523835-2026-00628
Device Sequence Number21107753
Product Code MSS
UDI-Device Identifier00380659777745
UDI-Public(01)00380659777745(17)320131(11)220127(10)147LK8
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K212039
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 06/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIV
Device Catalogue Number8065977774
Device Lot Number147LK8
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/01/2026
Initial Report FDA Received Date06/25/2026
Date Device Manufactured01/27/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CLAREON TORIC IOL; MONARCH III IOL, CARTRIDGE D; PROVISC OPHTHALMIC
Patient Age77 YR
Patient SexMale
-
-