Model Number 383069
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/28/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the low voltage portion of the right ventricular (rv) defibrillation lead was suspected to have fractured.The lead was explanted and replaced.It was further reported that the lead was returned to the manufacturer after being prophylactically replaced and subsequently tested out of specification during manufacturer's analysis.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the full lead was returned in segments, analyzed.Analysis indicated the outer insulation of the lead developed a breach due to metal ion oxidation while in vivo.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELECTSECURE MRI SURESCAN
Common Device Name
DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key25620440
Report Number2649622-2026-18933
Device Sequence Number10218170
Product Code NVN
UDI-Device Identifier00643169857681
UDI-Public(01)00643169857681(17)190626
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P030036
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date06/26/2019
Device Model Number383069
Device Catalogue Number383069
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2026
Initial Report FDA Received Date06/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/16/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DDMB1D4 ICD, 6935M62 LEAD
Patient Age48 YR
Patient SexMale
-
-