RESPIRONICS, INC. TRILOGY; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number EU2110X15B
Medical Device Problem Codes Circuit Failure (1089); Device Sensing Problem (2917)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/19/2026
Type of Reportable Event Malfunction
Event or Problem Description
A trilogy evo device was returned to the third-party service center with an allegation of a ventilator inoperative alarm after a software update.No medical intervention was specified.The device was not reported to be in clinical use.The trilogy evo was returned to the third-party service center for evaluation, and the customer¿s complaint was confirmed.In addition, during the evaluation it was noted that a code regarding a failure in the machine flow sensor was observed.In conclusion, the device's pca (patient circuit assembly) and the machine flow sensor needs to be replaced to address the issue.Should further investigation provide new information that impacts the outcome of the investigation or the associated medical device reporting, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported failure of a ventilator inoperative alarm is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The reported failure of a machine flow sensor error is accommodated in the product risk management file as being linked to hazard with description patient exposure to function / deterioration of function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.Multiple service activities occur during the device¿s useful life according to the device¿s service manual.These service activities, in addition to any prior complaint investigations, have altered the device from its original state during manufacturing.Any issue found during manufacturing that is potentially documented in the dhr for this device does not have a causal relationship to the allegations or issues in this complaint.Therefore, the dhr for this device will not be reviewed at this time.However, the dhr files for this device are available and can be accessed upon specific request.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key25620982
Report Number2518422-2026-025988
Device Sequence Number21078702
Product Code CBK
UDI-Device Identifier00606959054981
UDI-Public(01)00606959054981
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberEU2110X15B
Device Catalogue NumberEU2110X15B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2026
Initial Report FDA Received Date06/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-