| Catalog Number |
TRM-240 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: rupture, capsular contracture baker grade unknown.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture" and "rupture".This record is for left side.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture baker grade iv" and "rupture".This record is for left side.The device was explanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d.9, h.3, h.6.Device evaluation: per the investigation procedure, the device is analyzed through visual microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
- rupture: observed an opening through microscopic inspection assessed as unidentified (tear) opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, b6, d6b, h6.
|
| |
|