| Model Number |
URF-P7R |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Based on the results of the investigation, it is likely the following led to the malfunction: cause traced to failure of the adhesive.Date of event is unknown.Additional information will be provided once available.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the uretero-reno fiberscope to the service center for a separate issue.During the device analysis, the service repair technician indicates that the bending section cover glue was missing.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|