|
A 13 months retrospective review revealed 9 occurrences of complaints related to discrepant hba1c result on the g8 analyzer.Data assessment indicated the reported events were mostly due to operational problem and/or patient sample problem, obstruction of flow, maintenance problem and mechanical failure.The results were improved either by the customer rerunning affected samples, adjusting retention time where applicable samples or replacing failed parts.The g8 analyzer functioned as designed by alerting the operator with error flags when applicable.There was no indication of incorrect patient results being generated due to fault or failure of the operation of the g8 analyzer that requires further escalation.The g8 variant analysis mode operator¿s manual v 3.4.2 under chapter 1: introduction and applications: specimen collection and handling: collect venous whole blood specimens in vacuum collection tubes containing k2- edta or k3-edta and mix thoroughly.Specimens may be stored up to fourteen days at 2-8°c before analysis.Specimens may be stored up to twenty-four hours at room temperature (10-25ºc) before analysis.The minimum volume required for analysis directly from collection tubes is 1 ml of whole blood.Venous whole blood samples as small as 50 l may be used when appropriate sample cup and software options are selected.Quality control: in order to monitor and evaluate the accuracy and precision of the analytical performance, controls should be assayed daily and after column replacement.Tosoh suggests running at least two levels of quality control material.The mean of one should be in the non-diabetic range (4-7% hba1c) with the second in the range of 9-12% hba1c.If the value of one or more control specimens is out of the acceptable range, recalibrate the system and rerun the controls before testing patient samples.Qc materials should be used in accordance with local, state, federal and accredited organizations.Controls should be diluted with hsi hemolysis & wash solution to obtain an optimal total area (ta) in the range of 700-3000.However, a ta in the range of 500-4000 is acceptable and reportable for whole blood specimens.Limitations of the procedure: abnormal red cell survival the life span of red blood cells is shortened in patients with hemolytic anemias, and the actual life span depends upon the severity of the anemia.As a consequence, specimens from such patients may exhibit decreased glycohemoglobin levels compared to patients with normal red cell life span.The life span of red blood cells is lengthened in polycythemia or post-splenectomy patients.Specimens from such patients may exhibit increased glycohemoglobin levels.The most probable cause of the reported event could not be established.
|